GU60A Ceiling Height Requirement« Back to Questions List

What is the ceiling height requirement for the GU60A?
Posted by Medical Imaging
Asked on April 15, 2015 1:59 pm

For the GU60, please keep in mind for the ceiling height they show the minimum height of 2.8m. This will result in some loss of functionality. The ideal ceiling height for the GU60 is 10’ 4”. 10’ will work fine with one restriction. Max SID and max arm height will not allow complete CCW rotation (this would not be a normal function). 9’ 4” to 9’ 10” will work fine but will impact potential table position height (Max height of 39” is achieved with a 10’ ceiling). See full layout guidelines are in the Pre-Installation appendix and also page 21 of the Installation Manual.

Posted by samsungDRadmin
Answered On April 15, 2015 2:15 pm